FDA RECALL RECORD · Z-0480-2026

GM85 Digital Mobile X-ray imaging System; Recall

GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

FDA recall Class II

Ongoing

FDA lists this Class II recall for GM85 Digital Mobile X-ray imaging System;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0480-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 7, 2025
FDA report date
Nov 19, 2025
Quantity in commerce
236 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Affected model & UDI checker

Search the exact identifiers published for Z-0480-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
GM85GM85 Digital Mobile X-ray imaging System;Check source noticeNot specified
Not specifiedGM85 Digital Mobile X-ray imaging System;Check source notice08806088582412
FDA code information

Model Number: GM85; UDI-DI: 08806088582412; Serial Numbers: 514GM3IK400001L, 5143M3IK300001X, 5143M3IJ100010M, 5143M3IHC00014H, 5143M3IK300008R, 5143M3JJ400007E, 5143M3HJ900002V, 5143M3IK400004F, 513YM3IK500003Z, 5143M3HK600005V, 5143M3JJ200010J, 5143M3HK600008M, 5143M3HK200004H, 5143M3JK500003W, 5143M3IJ200002F, 5143M3IHC00007A, 5143M3JJ300001K, 5143M3IJ600014H, 5143M3JK500006X, 5143M3JJ200017Y, 5143M3JK400014E, 5143M3IJ600013B, 5143M3GK500004A, 5143M3IHC00011V, 5143M3JK500001M, 5143M3JJ300012N, 5143M3GK500005M, 5143M3IK400007W, 5143M3JK400008N, 5143M3JJ300011Z, 5143M3JK100003Z, 5143M3JK500007L, 5143M3IK500001F, 5143M3IHC00002L, 5143M3IK300006T, 5143M3IHC00010R, 5143M3IK600004K, 5143M3IJ300003E, 5143M3IHC00003Y, 5143M3JJ400006K, 5143M3HJ900001L, 5143M3IJ600003L, 5143M3IJC00001Y, 5143M3IJ600001H, 5143M3HJ500003A, 5149M3IK400003E, 5143M3IJ800008L, 5143M3JJ700002R, 5143M3HK300002V, 5143M3JJ200008B, 513YM3IK500002R, 5143M3JK100005D, 5143M3IJ600004F, 5143M3IHC00005J, 5143M3JJC00004Y, 5143M3HK600003R, 5143M3JK500004D, 5143M3HK600004A, 5143M3HK300012L, 5143M3HK300008M, 5143M3JJ200015M, 5143M3HK300007N, 5143M3JK100007F, 5143M3IHC00001H, 5143M3IJ400006V, 5143M3IJ100002P, 514GM3HK100003J, 5143M3HK600010W, 5143M3HJ500001Y, 5143M3JK400010W, 5143M3IJ300005Z, 5143M3HK400010M, 5149M3IK400002K, 5143M3IK400002J, 5143M3GK500007W, 5143M3IK400009H, 5143M3IJ300002Y, 5143M3JK100004A, 513YM3IK400001W, 5143M3HK600011N, 5143M3HK300003Y, 5143M3JJ200012Z, 5143M3JK400007B, 5143M3HJA00002K, 5149M3IK400001H, 5143M3IJ100005R, 5143M3IHC00012Z, 5143M3HK400009W, 514GM3HK100002R, 5143M3IJ100001Y, 5143M3HK700002K, 5143M3IJ400004B, 5143M3JK400004M, 5143M3HJ900006H, 5143M3HK300005W, 5143M3JJ700001V, 5143M3IJ600015P, 5143M3JJ500002M, 5143M3HJ900003W, 5143M3IK600001Y, 5143M3JK100015B, 5143M3HJ600003K, 5143M3IJ200001D, 5143M3RJ100001M, 5143M3HK600009X, 5143M3IK600003N, 5143M3JJ300007T, 5143M3IJ400003R, 5143M3IK600002W, 5143M3HK200006K, 5143M3HJ900007R, 5143M3JK100002B, 5143M3IJ300006N, 5143M3IHC00017N, 5143M3IK300009L, 5143M3IJ500001B, 5143M3JJC00003L, 5143M3HJ500004M, 5143M3JK100006N, 5143M3HK700001D, 514GM3JK300001A, 5143M3IJ800005X, 5143M3HK400003P, 5143M3IK400005T, 5143M3HK300009K, 5143M3IJ600012X, 5143M3IHC00015L, 5143M3HK300010P, 5143M3RJ100003T, 5143M3DHB00001M, 5143M3IJ100003X, 5143M3IJ600006W, 5143M3IK400008A, 5143M3IHC00019T, 5143M3JK100014L, 513YM3IK400002A, 513YM3IK500005Y, 5143M3JK500002P, 5143M3HK400004J, 5143M3HJ500002J, 5143M3IJ400005N, 5143M3JJ400002X, 5143M3JJ200014F, 5143M3JJ700004K, 5143M3HK300017E, 5143M3IJ800011E, 5143M3HK300013A, 5143M3HK600007Z, 5143M3HK300004E, 5143M3GK500003J, 5143M3GK500001H, 5143M3GK500002Y, 5143M3GK500006P, 5143M3HK600002T, 5143M3HK600001D, 5143M3JK400013K, 5143M3HK400006F, 5143M3HK400007T, 5143M3HK400005Z, 5143M3HK400008X, 5143M3JK500005V, 513YM3IK500001K, 513YM3IK500004H, 5143M3HK300019Z, 5143M3JK400011A, 5143M3JK400003D, 5143M3JK400006R, 5143M3JK400009V, 5143M3JK400012H, 5143M3IK400003Z, 5143M3IK400006X, 5143M3HK300016Y, 5143M3IK300005H, 5143M3IK300007P, 5143M3HK300006Z, 5143M3HK300001F, 5143M3IK300002J, 5143M3HK200005V, 513YM3IK100002D, 513YM3IK100003N, 5143M3JK100013W, 514GM3HK100001E, 514GM3HK100004W, 5143M3JK100009M, 5143M3JK100008H, 5143M3JJC00006J, 5143M3JJC00002H, 5143M3HJA00010W, 5143M3HJA00008X, 513YM3IJA00001X, 5143M3HJA00007D, 5143M3HJA00001B, 5143M3HJ900010Z, 5143M3HJ900009E, 5143M3IJ800012D, 5143M3IJ800014B, 5143M3IJ800009M, 5143M3IJ800013Y, 514GM3JJ900001R, 5143M3IJ800007K, 5143M3IJ800001B, 5143M3IJ800002V, 5143M3IJ800003J, 5143M3IJ800004F, 5143M3JJ700003D, 5143M3JJ700005W, 5143M3HJ600005D, 5143M3IJ600009J, 5143M3IJ600008K, 5143M3IJ600007N, 5143M3IJ600005Y, 5143M3IJ600011M, 5143M3HJ500006W, 5143M3IJ600002P, 5143M3IJ400007P, 5143M3JJ400005H, 5143M3IJ300004W, 5143M3JJ300009R, 5143M3JJ300004B, 5143M3JJ300005D, 5143M3JJ200018X, 5143M3JJ300002X, 5143M3JJ300003J, 5143M3JJ200013D, 5143M3JJ200011L, 5143M3JJ200002M, 5143M3JJ200005X, 5143M3JJ200009A, 5143M3JJ200006R, 5143M3RJ100002V, 5143M3IJ100011V, 5143M3IJ100004E, 5143M3IHC00016Y, 5143M3IHC00009M, 5143M3IHC00008E, 5143M3IHC00006T;

View the original FDA recall notice

Why was this product recalled?

The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

What actions did NeuroLogica Corporation take?

On October 7, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. The letters described the problem, product and actions to be taken. The customers were instructed to take the following Actions: Users may continue to use the GM85 system. Customers can inspect the unit for missing or loose bolts and if any issues with the bolts are detected, then please contact the service engineer immediately for inspection. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Complete and return the Customer Reply Form (CRF) Samsung Electronics will undertake the voluntary corrective action to reinforce the arm with a high tensile bolt to eliminate the risk of the THU falling and preventing any potential risk of injury. Service engineers will correct all affected devices free of charge and contact you to arrange for the correction. All corrections in the field started in September 2025 and will be completed by June 2026. If you have any questions, contact VP via email: ngujar@neurologica.com or call 978-564-8632.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology