Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0481-2026

SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600 Recall

SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600

FDA recall Class II

Ongoing

FDA lists this Class II recall for SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0481-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 7, 2025
FDA report date
Nov 19, 2025
Quantity in commerce
56 units (55 US, 1 OUS)
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600

Affected model & UDI checker

Search the exact identifiers published for Z-0481-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600Check source notice00606959032385
FDA code information

Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;

View the original FDA recall notice

Why was this product recalled?

Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.

What actions did Philips Respironics, Inc. take?

On October 7, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions for Distributors: Check inventory and place into quarantine any identified affected products. Complete and return the provided response form. Alert any customers who may have purchased affected products and work with them to return the affected product. A Philips representative will reach out regarding return and replacement of all affected product. Actions for patients: Check the serial numbers to verify if product affected by recall. If your product is affected, please discontinue use or the battery immediately. Contact your distributor and follow their instructions on how to receive a replacement battery and arrange for the return of the affected batteries in your possession. If the serial number on your battery kit does not match any of the affected serial numbers, then your battery is not affected by this issue and there is no need for any further action by you.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.