Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0485-2026

Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT; Recall

Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0485-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 25, 2025
FDA report date
Nov 19, 2025
Quantity in commerce
3263 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

CFN: SIGSDL45CTVT; UDI-DIs: (1) 20884521741840, (2) 10884521741843; Lot numbers: N5F1971Y, N5F1723Y, N5D1858UY, N5B1134UY;

View the original FDA recall notice

Why was this product recalled?

A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Covidien LP take?

On September 25, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Customer actions: Identify and quarantine all unused impacted product listed. Return all quarantined devices to Medtronic. If purchased from a distributor, contact your distributor directly to arrange for the return of the devices back to your distributor. Please complete and return the enclosed Customer Confirmation Form. Pass on this notice to all those who need to be aware within your organization or to any organization where the listed affected lots may have been transferred or distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.