FDA RECALL RECORD · Z-0504-2025

Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victims of sudden cardiac arrest (SCA.) Recall

Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victims of sudden cardiac arrest (SCA.)

FDA recall Class II

Ongoing

FDA lists this Class II recall for Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victims of sudden cardiac arrest (SCA.). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0504-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 14, 2024
FDA report date
Dec 4, 2024
Quantity in commerce
22,294 total
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victims of sudden cardiac arrest (SCA.)

Affected model & UDI checker

Search the exact identifiers published for Z-0504-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedDefibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victims of sudden cardiac arrest (SCA.)Check source notice00815098020447
FDA code information

UDI-DI: 00815098020447, DDP-2001 adult defibrillation pads, single pack; 30815098020448, DDP-2001 adult defibrillation pads, 10 pairs; 50815098020442, DDP-2001 adult defibrillation pads, 40 pairs

View the original FDA recall notice

Why was this product recalled?

Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Defibtech, LLC take?

Defibtech issued Urgent Field Safety Notice on 10/14/24 by e-mail to the Irish Market h via the Distributor Oxygen Care Ltd. titled: Unauthorized Label Applied to Defibtech, LLC's Automated External Defibrillator (AED} Pads within the Irish Market. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/User: 1. Check the label of your Defibtech AED Pads. Please review Attachment B which clearly shows authorised labelled Defibtech AED Pads versus unauthorised labelled Defibtech AED Pads. 2. Please confirm receipt of this notice by completing Attachment C "Customer Reply Form", attached to this notice, and return to Oxygen Care, Ltd (our new official Irish Distributor) so that our records may be updated. 3. If an unauthorised label applied to Defibtech AED Pads is confirmed, please urgently return the Defibtech AED Pads along with the completed "Customer Reply Form" to Oxygen Care. Upon receipt of the unauthorized labelled Defibtech AED Pads, Oxygen Care will send a replacement set of authorised labelled Defibtech AED Pads to you at no charge: Oxygen Care Ltd.: Email: Sales@ oxyge n-care.ie Phone: 01 2769700 Address: 2 Holfeld Business Park, Kilmacanogue. Co. Wicklow. Eircode A98 NP27 4. If you have transferred the Defibtech AED Pads to another entity or department, please contact Oxygen Care as per the above contact details. The unauthorised labelled Defibtech AED pads were re-labelled with expiration dates beyond the shelf life of the Defibtech AED Pads. The affected Defibtech AED Pads should not be used. Defibtech will continue to work closely with regulatory authorities toward the resolution of this issue. Defibtech is committed to patient safety and is focused on providing the highest level of customer satisfaction. We appreciate your cooperation and sincerely apologize for any inconvenience this may cause you.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.