Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL Recall
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
FDA recall Class II
Ongoing
FDA lists this Class II recall for Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0554-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 9, 2024
- FDA report date
- Dec 4, 2024
- Quantity in commerce
- 48
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Affected model & UDI checker
Search the exact identifiers published for Z-0554-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| SIL-40C | Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL | Check source notice | Not specified |
FDA code information
Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.
Why was this product recalled?
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Shimadzu Corp. Analytical & Measuring Instruments Div. take?
On 10/9/24, field safety notices were mailed to customers asking them to do the following: Check the Serial Numbers, and if the instruments are affected by this issue, contact your regional office. The corrective actions to be taken to affective products are (1) Add a cover part to the main board. (2) Update the SIL-40 firmware to Ver. 1.16 or above. For questions, please call 800-477-1227 between the hours of 8:30-5:00 EST or visit www.ssi.shimadzu.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0554-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology