FDA RECALL RECORD · Z-0558-2026

SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 L; Prepared with Balanced Heparin Recall

SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 ¿L; Prepared with Balanced Heparin, Model/Catalog Number: 06 0187 Product Description: A clear, plastic capillary blood collection tube

FDA recall Class II

Ongoing

FDA lists this Class II recall for SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 L; Prepared with Balanced Heparin. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0558-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 26, 2025
FDA report date
Nov 26, 2025
Quantity in commerce
352750
Distribution
See key details
FDA cause
Process change control

What is being recalled?

SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 L; Prepared with Balanced Heparin, Model/Catalog Number: 06 0187 Product Description: A clear, plastic capillary blood collection tube

Affected model & UDI checker

Search the exact identifiers published for Z-0558-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedSAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 L; Prepared with Balanced HeparinCheck source notice10643351000281
FDA code information

Model Number: 06 0187 GTIN (Inner Pack): 30643351000285 UDI-DI: 10643351000281 All Lots produced from 31JUL2023 to 28FEB2025: 23I4063, 24A4108, 24B4197, 24D4023, 24E4137, 24E4278, 24L4004

View the original FDA recall notice

Why was this product recalled?

Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.

What actions did ASP Global, LLC. dba Anatomy Supply Partners, LLC. take?

On 09/26-29/2025, ASP Global emailed an URGENT MEDICAL DEVICE RECALL Letter to customers. ASP Global, LLC, is voluntarily recalling all RAM Scientific SAFE-T-FILL products within remaining expiry. ASP Global was made aware that our contract manufacturer for the RAM Scientific SAFE-T-FILL products was issued an FDA Warning Letter citing significant violations of the Quality System requirements, including inadequate production and process controls, as well as design controls. Customers are instructed to: a) Discontinue use of the recalled products immediately. b) Quarantine any remaining inventory to prevent further use. c) If critical or abnormal test results were obtained using the affected SAFE-T-FILL products, ensure they are verified through repeat or confirmatory testing in accordance with institutional protocols (e.g., venous confirmation of abnormal pediatric screening results). d) Dispose remaining inventory in accordance with your local facility guidelines or notify ASP Global at FieldActions@aspglobal.com for return authorization and disposal. e) Complete and return the Recall Response Form to confirm receipt and actions to FieldActions@aspglobal.com. For questions or assistance - contact FieldActions@aspglobal.com (855)-867-2190| Mon Fri, 9:00 AM 5:00 PM ET

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology