Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay Recall
Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay
FDA recall Class II
Ongoing
FDA lists this Class II recall for Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0569-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 7, 2025
- FDA report date
- Nov 26, 2025
- Quantity in commerce
- 109 in total
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay
Affected model & UDI checker
Search the exact identifiers published for Z-0569-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 480050 | Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay | Check source notice | Not specified |
| Not specified | Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay | Check source notice | 00840733102196 |
FDA code information
UDI/DI 00840733102196, Lot Numbers: 480350U020, Exp. 2026-06-12
Why was this product recalled?
The cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Meridian Bioscience Inc take?
Meridian Bioscience issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees beginning on 10/15/2025 via email. The notice explained the problem with the device, health hazard, and requested the following: "Required Actions: " Identify remaining inventory of Alethia C. difficile, Lot 480050U036. " Identify the Reaction Buffer bag located within the kit box. " Remove Reaction Buffer Tubes from Bag and inspect each cap for the presence of a gasket. " Segregate Reaction Buffer Tubes which have no gasket present. Record the quantity of tubes with no gasket present on the CONFIRMATION OF NOTIFICATION Form. " Discard Reaction Buffer Tubes which have no gasket present in accordance with local, state, or federal regulations. Note: Execution of the test procedure requires the use of a Reaction Buffer tube. The remaining components in this kit may continue to be utilized for patient testing so long as a new Reaction Buffer tube is available. Once the Reaction Buffer is depleted, the remaining kit components must be discarded in accordance with local regulations. " Meridian Bioscience Inc. will provide credit for the number of tests impacted. " Complete and return the CONFIRMATION OF NOTIFICATION Form to: Product Support Manager, Meridian Bioscience, Inc., 3471 River Hills Drive, Cincinnati, OH 45244; Fax: (513) 272-5432 or Email: FieldActionSupport@meridianbioscience.com. Complete this form if there is no remaining inventory." For any question regarding this notification please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at FieldActionSupport@meridianbioscience.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0569-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology