Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Recall
Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an Olympus endoscope to deliver a nylon loop snare.
FDA recall Class I
Ongoing
FDA lists this Class I recall for Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0570-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Oct 30, 2025
- FDA report date
- Dec 3, 2025
- Quantity in commerce
- 7803 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an Olympus endoscope to deliver a nylon loop snare.
Affected model & UDI checker
Search the exact identifiers published for Z-0570-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| HX-400U-30 | Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 | Check source notice | 04953170368615 |
FDA code information
Model/Catalog Number: HX-400U-30; UDI: 04953170368615; All Lots which have not expired;
Why was this product recalled?
Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Olympus Corporation of the Americas take?
On October 30, 2025 and again on November 3, 2025, URGENT: MEDICAL DEVICE CORRECTIVE ACTION letters were sent to customers. Actions to be taken: Users are reminded of the importance of adhering to the Warnings and Cautions statements that are present in the instructions for use (IFU, ref: GK4574 Rev 16). A diagram was provided to illustrate the need to maintain a straight position of the proximal end of the tube sheath to prevent increased frictional resistance which may lead to breakage of the operating piipe when the slider is pushed. Particular attention should be paid to the following key guidance within Sections 10 and 12 of the IFU. Carefully read the content of this letter, including the supplemental guidance and reinforced text from the IFU as noted above. a) Always have Olympus loop cutter available when using the PolyLoop device b) When operating the yellow tube joint (yellow cylinder), be careful not to move it away from the handle, prematurely tightening the loop which may cause the inability to release the ligation loop. c) If resistance is encountered, do not continue or force the advancement of the slider located on the device handle. 2. Ensure all personnel are completely knowledgeable on the content of this notification. You may continue to use the device and are reminded of the importance of adhering to the warnings that are present in the instructions for use. 3. Olympus requests that you acknowledge receipt of this letter through our recall web portal. 4. If you have further distributed this product, identify your customers, and forward this notification to them. ***Update 04/24/2026*** On April 20, 2026 URGENT: MEDICAL DEVICE CORRECTIVE ACTION UPDATE letters were sent to customers. Olympus has updated the Instructions for Use (IFU) for the Single Use Ligating Device to clarify correct handling and loop-release techniques. The revised IFU includes previously issued guidance as well as supplemental explanations and illustrations intended to clari
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0570-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology