TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509) Recall
TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)
FDA recall Class II
Ongoing
FDA lists this Class II recall for TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0572-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 11, 2024
- FDA report date
- Dec 4, 2024
- Quantity in commerce
- 160 units
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509)
Affected model & UDI checker
Search the exact identifiers published for Z-0572-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 5161-00-5048 | TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509) | Check source notice | Not specified |
| Not specified | TriALTIS TORQUE LIMITER, RATCHET, 11.3 Nm, Model Number 5161-00-5048 (G919509) | Check source notice | 00858325005688 |
FDA code information
Model No 5161-00-5048; UDI-DI 00858325005688; Lot Numbers: GB153869 and GB156110
Why was this product recalled?
An out of specification component may result in the device exhibiting torque output higher than the design specification.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Gauthier Biomedical, Inc. take?
A Medical Device Recall (Removal) notification letter dated 11/11/24 was sent to the customer. Please Take the Following Steps: 1. Examine your inventory to determine if you have the affected lots and quarantine them immediately. DO NOT USE THE AFFECTED INSTRUMENTS. 2. Follow standard product return practices by contacting Heidi Wright at heidi.wright@gauthierbiomedical.com to issue an RMA number and coordinate the return of the affected product(s) to the following address: ATTN: Heidi Wright Gauthier Biomedical, Inc. 2221 Washington Street Grafton, WI 53024 Note: If the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual on this page of the notification. 3. Review, complete, sign and return the attached Verification Form to Heidi Wright at heidi.wright@gauthierbiomedical.com within five (5) business days of receipt of this notification. Please include in the email subject: CA 24-010. 4. Please complete the attached Verification Form even if you do not have the affected product lots on hand. 5. Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage transport, store, stock, or use the affected products). 6. If any of the affected product lots have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness and keep a copy for your records. 8. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting program online: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reportingprogram/reporting-serious-problems-fda. This medical device recall (removal) has been reported to the relevant health authorities. If you have any questions, please contact Heidi Wright, Quality Manager at Gauth
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0572-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology