Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0572-2026

Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Recall

Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0572-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 17, 2025
FDA report date
Nov 26, 2025
Quantity in commerce
60 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length

Affected model & UDI checker

Search the exact identifiers published for Z-0572-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
BPCN0815KBrand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon CatheterCheck source notice00821925033016
FDA code information

Model: BPCN0815K; UDI: 00821925033016; Lot Number: 25037006;

View the original FDA recall notice

Why was this product recalled?

Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.

What actions did Olympus Corporation of the Americas take?

On October 17, 2025, "URGENT: MEDICAL DEVICE REMOVAL" letters were sent to customers. Actions required: 1. 1. Examine your inventory and quarantine any affected devices. 2. Cease usage of the product with immediate effect. 3. Return any remaining affected product, please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will issue a credit to your facility upon return of your affected product. 4. Please forward this notice to other users who may have the affected products if you have further distributed it. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA If you require additional information, please do not hesitate to contact directly by phone at (647) 999-3203 or by e-mail Cynthia.Ow@Olympus.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.