FDA RECALL RECORD · Z-0579-2026

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Recall

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0579-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 14, 2025
FDA report date
Dec 3, 2025
Quantity in commerce
7023 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Affected model & UDI checker

Search the exact identifiers published for Z-0579-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
TJF-Q190VBrand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190VCheck source notice04953170405563
Not specifiedBrand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190VCheck source notice04953170452024
FDA code information

Model/Catalog Number: TJF-Q190V; UDI: 04953170405563 / 04953170452024; Serial Numbers: All;

View the original FDA recall notice

Why was this product recalled?

Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Olympus Corporation of the Americas take?

On October 14, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Carefulle read the notification and ensure that all personnel are completely knowledgeable and thoroughly trained on the content of the letter. Complete the required training provided in the letter. Keep a copy of the notification with the operation manual. It is recommended to receive on-site TJF duodenoscope reprocessing training at your facility. If you have further distributed this product, identify and forward them this notification.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology