ErgoStar CM 40 Recall
ErgoStar CM 40 Model/Catalog Number: MP01840 Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.
FDA recall Class I
Ongoing
FDA lists this Class I recall for ErgoStar CM 40. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0590-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Nov 10, 2025
- FDA report date
- Dec 17, 2025
- Quantity in commerce
- 14420 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
ErgoStar CM 40 Model/Catalog Number: MP01840 Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.
Affected model & UDI checker
Search the exact identifiers published for Z-0590-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MP01840 | ErgoStar CM 40 | Check source notice | Not specified |
| Not specified | ErgoStar CM 40 | Check source notice | 04048675420736 |
FDA code information
Model/Catalog Number: MP01840; UDI Number:04048675420736; Lot numbers: All lot numbers
Why was this product recalled?
Multiple complaints were reported in which cracks formed in the hose of specific catheter mounts.
What actions did Draeger, Inc. take?
On November 10, 2025, Urgent Medical Device Recall letters were sent to customers. Necessary Action: Identify if you have any affected product. Do not use any unused ErgoStar CM 40/CM 45/CM 55/CM 60 in your stock. Unused stock should be removed and returned to Draeger. Draeger recommends the alternative product MP01850 ErgoStar CM 50 as a suitable replacement. Please ensure that all potential users in your facility are made aware of this Urgent Medical Device Recall Notice and the information contained herein. If you have forwarded affected products to third parties, please also forward this Urgent Medical Device Recall notification.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0590-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology