Brand Name: Galaxy System Product Name: Galaxy System Recall
Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedure.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Galaxy System Product Name: Galaxy System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0598-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 12, 2025
- FDA report date
- Dec 3, 2025
- Quantity in commerce
- 47
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedure.
Affected model & UDI checker
Search the exact identifiers published for Z-0598-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| EOQ | Brand Name: Galaxy System Product Name: Galaxy System | Check source notice | 00850048825048 |
| GAL-001 | Brand Name: Galaxy System Product Name: Galaxy System | Check source notice | Not specified |
FDA code information
Lot Code: Product Code: EOQ UDI: 00850048825048 SN: included in Customer Notification Letter (template) CRB-1502 CRB-1530 CRB-1558 CRB-1503 CRB-1531 CRB-1560 CRB-1508 CRB-1534 CRB-1562 CRB-1513 CRB-1535 CRB-1563 CRB-1514 CRB-1537 CRB-1564 CRB-1515 CRB-1539 CRB-1565 CRB-1516 CRB-1540 CRB-1566 CRB-1518 CRB-1541 CRB-1567 CRB-1519 CRB-1542 CRB-1568 CRB-1520 CRB-1545 CRB-1570 CRB-1521 CRB-1546 CRB-1571 CRB-1522 CRB-1553 CRB-1573 CRB-1523 CRB-1554 CRB-1574 CRB-1524 CRB-1555 CRB-1559 CRB-1525 CRB-1556 CRB-1569 CRB-1528 CRB-1557
Why was this product recalled?
Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.
What actions did Noah Medical Corporation take?
On 09/12/2025, the firm sent via FedEx an "Urgent: MEDICAL DEVICE RECALL" Letter to customers. The letter described the product, risk to health and action to be taken. Customers were notified that the password has been removed and they are instructed to: Log in and access the applicable IFU and all addendums. Noah Medical recommends that customers download the IFU and addendums for ease of reference. For questions contact: Noah Medical Customer Service email: customerservice@noahmed.com Noah hotline: 888-325-NOAH (1-888-325-6624)
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0598-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology