NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK; Recall
NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK;
FDA recall Class II
Ongoing
FDA lists this Class II recall for NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0601-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 31, 2025
- FDA report date
- Dec 3, 2025
- Quantity in commerce
- 39,148
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK;
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Product Number/UDI-DI (GTIN)/Lot Numbers: NN-8024 00816744028244 37AI2944 37AJ2013 37BD3753 37BD3778 37CA1605 37CB3738 37CH0636 37CI1919 37EH0405 37FC3059 37FD0154 37GA4512 37GH1715 37GI0110 37GJ1623 37HB0029 37HH2313 37HI1536 37HI1631 37IA7694 37ID4162 37IG2900 37JB1413 37JH0131 37JH2903 37KG1406 37KH2953 37LA0708 37LG1834; NN-8025 00816744028251 37BB2326 37BJ1836 37CH0640 37DC0545 37HD2766 37KB3677 37KJ1139; NN-8026 00816744028268 37BJ1837 37CC7852 37CH0643 37DJ1056; NN-8042 07613327663280 37CC9220 37ED9408 37FJ2202 37II2308 37LD8666
Why was this product recalled?
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Stryker Corporation take?
On October 31, 2025, the firm began notifying impacted customers through Urgent Medical Device Recall letters. Customers were instructed to mark affected product with "CONTAINS LATEX" and place a copy of the notice at the product's storage location before use. Product returns are not required. However, customers who would like to receive a replacement were given options to destroy or return product for a replacement.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0601-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology