Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0663-2025

Microscissor Tip (Disposable); Microline Catalog #: 3122. Recall

Microscissor Tip (Disposable); Microline Catalog #: 3122.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Microscissor Tip (Disposable); Microline Catalog #: 3122.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0663-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 5, 2024
FDA report date
Dec 18, 2024
Quantity in commerce
200 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Microscissor Tip (Disposable); Microline Catalog #: 3122.

Affected model & UDI checker

Search the exact identifiers published for Z-0663-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMicroscissor Tip (Disposable); Microline Catalog #: 3122.Check source notice811099010036
FDA code information

Microline Catalog #: 3122; UDI/DI: 811099010036; Lot Number: 00172110.

View the original FDA recall notice

Why was this product recalled?

A lot of ReNew Microscissor Tips exhibited a 2% rate of failure in terms of the heat shrink breaking off.

What actions did Microline Surgical, Inc. take?

On November 6, 2024 URGENT: MEDICAL DEVICE RECALL letters were sent to consignees. Actions to be taken by the Customer/User: Please locate those products you have that correspond with the above identified lots, quarantine them, and arrange for their return by contacting Microline Customer Service to receive an RMA number. Alternatively, you may wish to contact your Microline Distributor or Sales Representative to help with this. If you have used some or all of the identified product, please provide the item number, lot number, and the amount used so that we can track the progress of the recall. All products returned will be replaced at no cost. This recall is being done in an abundance of caution. No other item numbers or lots of ReNew Disposable Tips are affected by this recall. If you experience any adverse events or quality problems with the use of this product, please be sure to report this to the relevant Competent Authority s Adverse Event Reporting Program and to Microline as soon as possible. We regret any inconvenience and encourage you to contact Microline Customer Service at 978- 922-9810 or info@microlinesurgical.com. with any further questions. Actions to be taken by the Distributor: 1. Immediately determine if you are in possession of the products subject to this recall by checking the product name, item number, and lot number. 2. If you are a distributor of products subject to this recall, we require that you immediately remove affected products from inventory, quarantine products, and stop sales and shipping. 3. If you are a distributor and you have further distributed products subject to this recall, please identify customers immediately and notify them of this product recall using this Recall Notification Letter. Translation of this letter into local language is acceptable. We request that you direct your customers to return affected products to your facility for subsequent return to Microline. The relevant Competent Authority will be monitoring th

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology