Microscissor Tip (Disposable); Microline Catalog #: 3122. Recall
Microscissor Tip (Disposable); Microline Catalog #: 3122.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Microscissor Tip (Disposable); Microline Catalog #: 3122.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0663-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 5, 2024
- FDA report date
- Dec 18, 2024
- Quantity in commerce
- 200 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Microscissor Tip (Disposable); Microline Catalog #: 3122.
Affected model & UDI checker
Search the exact identifiers published for Z-0663-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Microscissor Tip (Disposable); Microline Catalog #: 3122. | Check source notice | 811099010036 |
FDA code information
Microline Catalog #: 3122; UDI/DI: 811099010036; Lot Number: 00172110.
Why was this product recalled?
A lot of ReNew Microscissor Tips exhibited a 2% rate of failure in terms of the heat shrink breaking off.
What actions did Microline Surgical, Inc. take?
On November 6, 2024 URGENT: MEDICAL DEVICE RECALL letters were sent to consignees. Actions to be taken by the Customer/User: Please locate those products you have that correspond with the above identified lots, quarantine them, and arrange for their return by contacting Microline Customer Service to receive an RMA number. Alternatively, you may wish to contact your Microline Distributor or Sales Representative to help with this. If you have used some or all of the identified product, please provide the item number, lot number, and the amount used so that we can track the progress of the recall. All products returned will be replaced at no cost. This recall is being done in an abundance of caution. No other item numbers or lots of ReNew Disposable Tips are affected by this recall. If you experience any adverse events or quality problems with the use of this product, please be sure to report this to the relevant Competent Authority s Adverse Event Reporting Program and to Microline as soon as possible. We regret any inconvenience and encourage you to contact Microline Customer Service at 978- 922-9810 or info@microlinesurgical.com. with any further questions. Actions to be taken by the Distributor: 1. Immediately determine if you are in possession of the products subject to this recall by checking the product name, item number, and lot number. 2. If you are a distributor of products subject to this recall, we require that you immediately remove affected products from inventory, quarantine products, and stop sales and shipping. 3. If you are a distributor and you have further distributed products subject to this recall, please identify customers immediately and notify them of this product recall using this Recall Notification Letter. Translation of this letter into local language is acceptable. We request that you direct your customers to return affected products to your facility for subsequent return to Microline. The relevant Competent Authority will be monitoring th
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0663-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology