Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0664-2025

MICRO-X Rover Mobile X-ray System, # MXU-RV35 Recall

MICRO-X Rover Mobile X-ray System, # MXU-RV35

FDA recall Class II

Ongoing

FDA lists this Class II recall for MICRO-X Rover Mobile X-ray System, # MXU-RV35. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0664-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 31, 2023
FDA report date
Dec 18, 2024
Quantity in commerce
5
Distribution
See key details
FDA cause
Software design

What is being recalled?

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Affected model & UDI checker

Search the exact identifiers published for Z-0664-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMICRO-X Rover Mobile X-ray System, # MXU-RV35Check source notice9357123000037
FDA code information

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

View the original FDA recall notice

Why was this product recalled?

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

What actions did Micro-X Ltd. take?

The firm will notify consignees using a service bulletin that was already posted. Additionally, the firm will implement a firmware update to extend the timer (4 ms to 10 ms minimum) to ensure even short exposures achieve the necessary image quality for diagnostic use.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.