Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0 Recall
Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0
FDA recall Class II
Ongoing
FDA lists this Class II recall for Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0678-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 31, 2024
- FDA report date
- Dec 18, 2024
- Quantity in commerce
- 778 systems
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0
Affected model & UDI checker
Search the exact identifiers published for Z-0678-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0 | Check source notice | 00859873006158 |
| Not specified | Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0 | Check source notice | 00859873006196 |
FDA code information
Hermes Medical Imaging System version 6.1 with Hybrid Recon version 4.0.x UDI-DI code: 00859873006158 Serial Numbers: Pending Hybrid Recon version 5.0.0 UDI code: (01)00859873006196(8012)005000000 Serial Numbers: Pending
Why was this product recalled?
Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Hermes Medical Solutions AB take?
On 10/31/2024, the firm emailed a "Field Safety Notice (FSN)" letter to customers informing them that when reconstructing a SPECT/CT study, in some cases, the SPECT and CT series are not correctly aligned. This may be due to separately performed acquisitions, or due to the camera operator re-setting the reference point (landmark) between SPECT and CT acquisitions. Normally, the Hybrid Recon application warns the user if the Frames of Reference do not match, however in some configurations this warning message may be turned off. Customers are instructed how to avoid the problem by: In all cases, the user must verify correct SPECT to CT alignment by careful inspection of the fused images during reconstruction. To ensure that you are warned about possible SPECT to CT misalignment, the Skip Frame of Reference warning radio button shall not be set on . The Skip Frame of Reference warning radio button can be found in the Program Parameters window (button at top right corner of the window left side of the ? symbol). Go to the Recon tab and find the Skip frame of reference warning radio button. If the button is on as shown here, this warning will not be shown to the users. The setting of the Skip Frame of Reference warning radio button should be checked to ensure it is off for each workflow Oncology, Neurology, Cardiology and Lung. Please contact your Hermes Medical Solutions support if you do not have permission to make the changes required. For questions or further assistance contact General: support@hermesmedical.com Canada: support.ca@hermesmedical.com USA: support.us@hermesmedical.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0678-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology