FDA RECALL RECORD · Z-0678-2025

Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0 Recall

Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0

FDA recall Class II

Ongoing

FDA lists this Class II recall for Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0678-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 31, 2024
FDA report date
Dec 18, 2024
Quantity in commerce
778 systems
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0

Affected model & UDI checker

Search the exact identifiers published for Z-0678-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedHybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0Check source notice00859873006158
Not specifiedHybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0Check source notice00859873006196
FDA code information

Hermes Medical Imaging System version 6.1 with Hybrid Recon version 4.0.x UDI-DI code: 00859873006158 Serial Numbers: Pending Hybrid Recon version 5.0.0 UDI code: (01)00859873006196(8012)005000000 Serial Numbers: Pending

View the original FDA recall notice

Why was this product recalled?

Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Hermes Medical Solutions AB take?

On 10/31/2024, the firm emailed a "Field Safety Notice (FSN)" letter to customers informing them that when reconstructing a SPECT/CT study, in some cases, the SPECT and CT series are not correctly aligned. This may be due to separately performed acquisitions, or due to the camera operator re-setting the reference point (landmark) between SPECT and CT acquisitions. Normally, the Hybrid Recon application warns the user if the Frames of Reference do not match, however in some configurations this warning message may be turned off. Customers are instructed how to avoid the problem by: In all cases, the user must verify correct SPECT to CT alignment by careful inspection of the fused images during reconstruction. To ensure that you are warned about possible SPECT to CT misalignment, the Skip Frame of Reference warning radio button shall not be set on . The Skip Frame of Reference warning radio button can be found in the Program Parameters window (button at top right corner of the window left side of the ? symbol). Go to the Recon tab and find the Skip frame of reference warning radio button. If the button is on as shown here, this warning will not be shown to the users. The setting of the Skip Frame of Reference warning radio button should be checked to ensure it is off for each workflow Oncology, Neurology, Cardiology and Lung. Please contact your Hermes Medical Solutions support if you do not have permission to make the changes required. For questions or further assistance contact General: support@hermesmedical.com Canada: support.ca@hermesmedical.com USA: support.us@hermesmedical.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.