Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0696-2025

HUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation device Recall

HUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation device

FDA recall Class II

Ongoing

FDA lists this Class II recall for HUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation device. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0696-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 18, 2024
FDA report date
Dec 25, 2024
Quantity in commerce
360 uinits
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

HUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation device

Affected model & UDI checker

Search the exact identifiers published for Z-0696-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
DYNJMBRC2BHUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation deviceCheck source noticeNot specified
Not specifiedHUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation deviceCheck source notice20193489005315
Not specifiedHUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tube fixation deviceCheck source notice10193489005318
FDA code information

UDI/DI 20193489005315 (case), 10193489005318 (each); All Lots

View the original FDA recall notice

Why was this product recalled?

Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

Medline Industries issued an URGENT MEDICAL DEVICE RECALL to its consignees of 11/18/2024 via USPS and email. The notice explained the issue, risk to user, and requested the destruction of the affected product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology