Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting. Recall

Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0712-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Dec 9, 2024
FDA report date
Jan 15, 2025
Quantity in commerce
11,101 units US
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.

Affected model & UDI checker

Search the exact identifiers published for Z-0712-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
VH-4001Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.Check source noticeNot specified
Not specifiedVasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.Check source notice00607567700451
FDA code information

Model Number: VH-4001. UDI-DI: 00607567700451. All unexpired Lots.

View the original FDA recall notice

Why was this product recalled?

Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Maquet Cardiovascular, LLC take?

Getinge notified consignees on 12/09/2024 via Urgent Medical Device Correction letter posted digitally on Getinge.com and sent to consignees via letter. Consignees were not instructed to return affected units, but instead to review the IFU information provided with the letter to minimize over-delivery of energy and mitigate risks. Getinge instructed consignees to notify customers if further distributed, as well as requested to complete and return the MEDICAL DEVICE CORRECTION RESPONSE FORM. Training to reinforce precautions and appropriate use of the device is expected to be deployed and communicated to all customers by end of January 2025. UPDATE: On 01/31/2025 an updated URGENT MEDICAL DEVICE CORRECTION letter was sent to customers. Consignees were instructed to review the IFU information provided in Appendix A of the letter and consider the listed areas to mitigate the risks. Getinge has created a training video, accessible via a QR code or link, to guide users on the proper use of the device and reinforce essential precautions. Consignees were instructed to notify customers if product was further distributed. Additionally, consignees were instructed to complete, sign and return the provided MEDICAL DEVICE CORRECTION RESPONSE FORM. UPDATE: On 07/11/2025 an updated Urgent Medical Device Correction Update letter was posted on Getinge.com and sent to consignees via letter. The updated communication informs users that the Instructions for Use (IFU) for the Vasoview Hemopro 2 EVH System has been revised to address the bent or detached heater wire and silicone peeling or detaching from the Jaws of the Harvesting Tool. Consignees were instructed to review the updated IFU, view the training video linked in the letter, complete and return the MEDICAL DEVICE CORRECTION RESPONSE FORM, and notify customers if product was further distributed.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology