Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev E_Vap. (Safety Filling adapter), Part Number: 115-066756-00. 2. Exchange Sev E_Vap. (Safety Filling adapter), Part Number: E115-066756-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery. Recall
Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev E_Vap. (Safety Filling adapter), Part Number: 115-066756-00. 2. Exchange Sev E_Vap. (Safety Filling adapter), Part Number: E115-066756-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev E_Vap. (Safety Filling adapter), Part Number: 115-066756-00. 2. Exchange Sev E_Vap. (Safety Filling adapter), Part Number: E115-066756-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0720-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 20, 2025
- FDA report date
- Dec 3, 2025
- Quantity in commerce
- 148 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev E_Vap. (Safety Filling adapter), Part Number: 115-066756-00. 2. Exchange Sev E_Vap. (Safety Filling adapter), Part Number: E115-066756-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery.
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
1. Part Number: 115-066756-00; DI: 06936415975777; Serial Numbers: AH6A45000871, AH6A45000880, AH6A45000881, AH6A46000929, AH6A46000931, AH6A46000944, AH6A47001200, AH6A44000657, AH6A44000658, AH6A44000661, AH6A44000664, AH6A44000677, AH6A44000678, AH6A44000680, AH6A44000768, AH6A44000778, AH6A46000925, AH6A46000927, AH6A46000928, AH6A46000935, AH6A46000937, AH6A46001114, AH6A47001159, AH6A47001164, AH6A47001174, AH6A47001201, AH6A45000872, AH6A46000932, AH6A46000939, AH6A46000940, AH6A47001196, AH6A47001202, AH6A47001207, AH6A47001215, AH6A46001102, AH6A46001111, AH6A46001116, AH6A47001160, AH6A47001161, AH6A47001163, AH6A47001165, AH6A47001166, AH6A47001167, AH6A47001178, AH6A46001097, AH6A46001117, AH6A44000645, AH6A44000647, AH6A44000648, AH6A44000651, AH6A44000655, AH6A44000656, AH6A44000659, AH6A44000660, AH6A44000662, AH6A44000684, AH6A44000686, AH6A44000687, AH6A44000688, AH6A44000690, AH6A44000691, AH6A44000695, AH6A44000696, AH6A44000698, AH6A44000700, AH6A44000702, AH6A44000703, AH6A44000770, AH6A44000776, AH6A46001090, AH6A46001109, AH6A46001118, AH6A47001171, AH6A47001176, AH6A44000771, AH6A44000772, AH6A45000815, AH6A45000819, AH6A45000821, AH6A45000822, AH6A45000824, AH6A45000825, AH6A45000829, AH6A45000832, AH6A45000834, AH6A45000835, AH6A45000868, AH6A45000869, AH6A45000870, AH6A45000873, AH6A45000874, AH6A45000875, AH6A45000876, AH6A45000877, AH6A45000878, AH6A45000882, AH6A45000883, AH6A45000884, AH6A46000926, AH6A46000930, AH6A46000936, AH6A46000943, AH6A44000779, AH6A44000674, AH6A44000692, AH6A44000777, AH6A45000879, AH6A47001170, AH6A44000765, AH6A44000769, AH6A47001175, AH6A28000535, AH6A46000918, AH6A46000923, AH6A46000924, AH6A46000934, AH6A46000938, AH6A46001091, AH6A46001093, AH6A46001099, AH6A46001100, AH6A46001101, AH6A46001103, AH6A46001104, AH6A46001107, AH6A46001108, AH6A46001110, AH6A46001113, AH6A47001177, AH6A44000679, AH6A44000683, AH6A44000689, AH6A44000697, AH6A44000649, AH6A44000650, AH6A44000652, AH6A44000676, AH6A44000681, AH6A44000682, AH6A44000699, AH6A44000704, AH6A44000766, AH6A44000773, AH6A44000775, AH6A44000780, AH6A47001216, AH6A44000693. 2. Part Number: E115-066756-00; DI: 06936415975777; Serial Number: AH6A45000885.
Why was this product recalled?
Potential for anesthesia leakage.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Mindray DS USA, Inc. dba Mindray North America take?
Mindray notified consignees on about 10/20/2025 via letter. Consignees were informed that they could continue using the units as normal, but if an anesthetic odor or leak is detected, contact Mindray Technical Support. Interventions may include switching to another V90 vaporizer or using intravenous anesthesia as clinically appropriate. For vaporizer replacement procedures, refer to A9 Operator s Manual (Part No. H-046-017199-00, Section 3.4 Vaporizer ), available on the Mindray website. Mindray s Technical Support team can assist and may be reached at (877) 913-9663 (Option 2) Monday through Friday, 8:30 a.m. 5:30 p.m. ET. Consignees were instructed to complete and return the Acknowledgement and Receipt form provided, ensure that all applicable clinical staff are aware of the issue and maintain a copy of the recall notification with affected units or IFUs. Additionally, they were instructed to notify any organization to which the affected V90 electronic vaporizers identified have been distributed. The Mindray Care Team will contact consignees to arrange the replacement of all affected SEVO and ISO V90 electronic vaporizers located at their facility. The V90 electronic vaporizers will be shipped directly to each facility, and a Mindray Care Team representative will schedule the installation and collect the affected V90 electronic vaporizers.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0720-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology