Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837 Recall
Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837
FDA recall Class II
Ongoing
FDA lists this Class II recall for Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0743-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 22, 2024
- FDA report date
- Jan 1, 2025
- Quantity in commerce
- 171 orders
- Distribution
- See key details
- FDA cause
- Software design (manufacturing process)
What is being recalled?
Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Only 3M Clarity Precision Grip Attachment Dual Arch Attachments are potentially affected. Part numbers/UDI-DI: 359-830/ 00652221307286; 359-831/ 00652221307293; 359-832/00652221307309; 359-833/ 00652221307316; 359-834/ 00652221307361; 359-835/ 00652221307378; 359-836/ 00652221307385; 359-837/ 00652221307392
Why was this product recalled?
Attachments may not mate with the tooth as intended in the digital treatment design.
What actions did 3M Company - Health Care Business take?
On November 22, 2024, the firm notified customers via URGENT Product Recall letters. Customers were informed that select grip attachments were impacted by a software nonconformance, limited to the lower arch of specific dual-arch Clarity Precision Grip Attachment orders. The firm advises that customers not use products from the identified Case IDs, and that they should dispose of affected products. Customers should complete and return the included customer product destruction form. The firm will provide replacement product. If you have additional questions or need to report an adverse event, contact the Clinician Support team (800-423-4588, option 2) or customortho@solventum.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0743-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology