Magnesium Flex reagent cartridge, 360 tests Recall
Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Magnesium Flex reagent cartridge, 360 tests. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0767-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 12, 2024
- FDA report date
- Jan 1, 2025
- Quantity in commerce
- 13,115 units (11,520 US, 1,595 ROW)
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;
Affected model & UDI checker
Search the exact identifiers published for Z-0767-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| K3057 | Magnesium Flex reagent cartridge, 360 tests | Check source notice | Not specified |
| Not specified | Magnesium Flex reagent cartridge, 360 tests | Check source notice | 00842768015694 |
FDA code information
Siemens Material Number (SMN): 10445158; Catalog Number: K3057; UDI-DI: 00842768015694; Lot numbers: 24057BB, 24064BC.
Why was this product recalled?
Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Siemens Healthcare Diagnostics, Inc. take?
On November 12, 2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Customer Actions Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Discontinue use of and discard the lots listed above. Review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within thirty (30) days. Please retain this letter with your laboratory records and forward this letter to those who may have received the product. Resolution Siemens Healthineers has introduced additional manufacturing process controls that have resolved this issue. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0767-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology