AURYON LASER SYSTEM 100-120 VAC. Recall
AURYON LASER SYSTEM 100-120 VAC. Model Number: EXM001. Product Number: EXM-2001-1100
FDA recall Class II
Ongoing
FDA lists this Class II recall for AURYON LASER SYSTEM 100-120 VAC.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0770-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 25, 2024
- FDA report date
- Jan 8, 2025
- Quantity in commerce
- 138 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
AURYON LASER SYSTEM 100-120 VAC. Model Number: EXM001. Product Number: EXM-2001-1100
Affected model & UDI checker
Search the exact identifiers published for Z-0770-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| EXM001 | AURYON LASER SYSTEM 100-120 VAC. | Check source notice | Not specified |
| Not specified | AURYON LASER SYSTEM 100-120 VAC. | Check source notice | 07290017590110 |
FDA code information
Product Number: EXM-2001-1100, Model Number: EXM001. UDI-DI: 07290017590110, Serial Number: EXM XXX.
Why was this product recalled?
Potential for procedural delays or interruptions during use of the Auryon Atherectomy System due to the Auryon Atherectomy System not advancing to the Activation (Ready) mode.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did AngioDynamics, Inc. take?
AngioDynamics notified consignees on about 11/25/2024 via URGENT VOLUNTARY MEDICAL DEVICE CORRECTION letter. Consignees were instructed to located affected devices on hand, place the revised Operator's Manual with affected systems, remove and destroy the Operator s Manual that is presently with the Auryon Laser System, and notify any customers of the recall if the Auryon Laser System has been redistributed to another facility. Additionally, consignees were asked to complete and return the Reply Verification Tracking Form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0770-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology