Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0770-2025

AURYON LASER SYSTEM 100-120 VAC. Recall

AURYON LASER SYSTEM 100-120 VAC. Model Number: EXM001. Product Number: EXM-2001-1100

FDA recall Class II

Ongoing

FDA lists this Class II recall for AURYON LASER SYSTEM 100-120 VAC.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0770-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 25, 2024
FDA report date
Jan 8, 2025
Quantity in commerce
138 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

AURYON LASER SYSTEM 100-120 VAC. Model Number: EXM001. Product Number: EXM-2001-1100

Affected model & UDI checker

Search the exact identifiers published for Z-0770-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
EXM001AURYON LASER SYSTEM 100-120 VAC.Check source noticeNot specified
Not specifiedAURYON LASER SYSTEM 100-120 VAC.Check source notice07290017590110
FDA code information

Product Number: EXM-2001-1100, Model Number: EXM001. UDI-DI: 07290017590110, Serial Number: EXM XXX.

View the original FDA recall notice

Why was this product recalled?

Potential for procedural delays or interruptions during use of the Auryon Atherectomy System due to the Auryon Atherectomy System not advancing to the Activation (Ready) mode.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AngioDynamics, Inc. take?

AngioDynamics notified consignees on about 11/25/2024 via URGENT VOLUNTARY MEDICAL DEVICE CORRECTION letter. Consignees were instructed to located affected devices on hand, place the revised Operator's Manual with affected systems, remove and destroy the Operator s Manual that is presently with the Auryon Laser System, and notify any customers of the recall if the Auryon Laser System has been redistributed to another facility. Additionally, consignees were asked to complete and return the Reply Verification Tracking Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology