SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C Recall
SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C
FDA recall Class II
Ongoing
FDA lists this Class II recall for SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0772-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 21, 2024
- FDA report date
- Jan 8, 2025
- Quantity in commerce
- 59 systems
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C
Affected model & UDI checker
Search the exact identifiers published for Z-0772-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 8637 | SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C | Check source notice | Not specified |
| 8667 | SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C | Check source notice | Not specified |
| Not specified | SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C | Check source notice | 00643169890961 |
FDA code information
GTIN: 00643169890961; Serial Numbers: NPL1000442 NPL1000448 NPL1000453 NPL1000765 NPL1000776 NPL1000797 NPL1000809 NPL1000810 NPL1000811 NPL1000813 NPL1000814 NPL1000815 NPL1000817 NPL1000818 NPL1000819 NPL1000827 NPL1000830 NPL1000834 NPL1000835 NPL1000858 NPL1001552 NPL1001558 NPL1001564 NPL1001566 NPL1001574 NPL1001583 NPL1001590 NPL1001591 NPL1001592 NPL1001594 NPL1001595 NPL1001601 NPL1001608 NPL1001615 NPL1001620 NPL1001629 NPL1001630 NPL1001631 NPL1001650 NPL1001651 NPL1001656 NPL1001670 NPL1001675 NPL1002257 NPL1002259 NPL1002260 NPL1002264 NPL1002265 NPL1002266 NPL1002267 NPL1002281 NPL1002282 NPL1002284 NPL1002286 NPL1002288 NPL1002295 NPL1002297 NPL1002308 NPL1002309
Why was this product recalled?
Software issues were identified in application version 2.x.
What actions did Medtronic Neuromodulation take?
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated November 2024 was sent to customers. Actions Share this notice with all those who need to be aware of this issue within your organization and maintain a copy of this notice in your records. Please complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information. The form may be emailed to neuro.quality@medtronic.com. Additional Information Medtronic is actively working on a permanent solution and will inform you as soon as it becomes available. Medtronic is communicating this information to the appropriate regulatory agencies. Adverse events or quality problems experienced with this product should be reported to Medtronic and the FDA MedWatch Adverse Event Reporting program via: Medtronic email at RS.NeuAdverseEventReporting@Medtronic.com or Medtronic Technical Services at 1-800-707-0933 weekdays 7:00 AM 6:00 PM CT. ***Update January 2026*** An URGENT: MEDICAL DEVICE CORRECTION notification letter dated January 2026 was sent to customers. The purpose of this letter is to inform you that a new version of the Model A810 SynchroMed Clinician Programmer Application (CP App), v2.0.3320, is now available for installation. This update addresses previously communicated software issues associated with earlier versions used for the Model 8637 SynchroMed II and Model 8667 SynchroMed III infusion pumps. These issues were initially communicated to you in the letter enclosed from November 2024. The new version of the A810 CP App resolves these issues. Customer Actions: " Download and install A810 CP App v2.0.3320 at your earliest convenience and contact your local Medtronic representative if you have any questions related to the update. " Sign and return the Confirmation Form enclosed with this letter, acknowledging that you have received this information. The form may be emailed to neuro.quality@medtronic.com. Retain a copy of this lette
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0772-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology