Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0795-2025

JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L Recall

JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L

FDA recall Class I

Ongoing

FDA lists this Class I recall for JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0795-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Dec 9, 2024
FDA report date
Jan 15, 2025
Quantity in commerce
6,520
Distribution
See key details
FDA cause
Component change control

What is being recalled?

JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L

Affected model & UDI checker

Search the exact identifiers published for Z-0795-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
O2C5LJMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5LCheck source noticeNot specified
Not specifiedJMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5LCheck source notice00092237623152
FDA code information

UDI-DI: 00092237623152. Serial Number Ranges: JA2311000001-JA2311000740, JA2311000741-JA2311001480, JA2311001481-JA2311002220, JA2311002221-JA2311002960, JA2311002961-JA2311003700, JA2312000001-JA2312000740, JA2312000741-JA2312001480, JA2312001481-JA2312002080, JA2401000001-JA2401000740.

View the original FDA recall notice

Why was this product recalled?

Oxygen concentrator may melt or catch on fire during use.

What actions did Jiangsu Jumao X-Care Medical Equipment Co., Ltd. take?

On 11/26/2024, a recall notice was sent a distributor asking them to take the steps below. On 12/09/2024, recall notices were sent from the distributor to customers who were asked to do the following: This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. All units from the specified serial number range MUST be immediately discontinued from use and removed from service and promptly returned. Complete and return the acknowledgement and receipt form via email to recall@compasshealthbrands.com If you have any questions, please contact: 1-800-376-7263 x444, Monday-Friday 8:00 am EST - 5:00 pm EST, or email recall@compasshealthbrands.com On 12/19/2024, additional recall notices were sent adding additional affected devices.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology