Ivalon Eye Wick 20cm- Intended to be used for the removal and collection of fluids from the Opthalmic surgical site. Recall
Ivalon Eye Wick 20cm- Intended to be used for the removal and collection of fluids from the Opthalmic surgical site. Model Number: Q604206
FDA recall Class II
Ongoing
FDA lists this Class II recall for Ivalon Eye Wick 20cm- Intended to be used for the removal and collection of fluids from the Opthalmic surgical site.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0798-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 22, 2024
- FDA report date
- Jan 8, 2025
- Quantity in commerce
- 120 units
- Distribution
- See key details
- FDA cause
- Equipment maintenance
What is being recalled?
Ivalon Eye Wick 20cm- Intended to be used for the removal and collection of fluids from the Opthalmic surgical site. Model Number: Q604206
Affected model & UDI checker
Search the exact identifiers published for Z-0798-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Q604206 | Ivalon Eye Wick 20cm- Intended to be used for the removal and collection of fluids from the Opthalmic surgical site. | Check source notice | Not specified |
FDA code information
LOT NUMBER: 24G1880
Why was this product recalled?
Potential for package seal to be out of specification and the sterile barrier cannot be guaranteed for the life of the product.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Carwild Corporation take?
Carwild Corp issued letter On Behalf Of First Aid Bandage Company (Fabco) Recall Notice Urgent via email 12/22/24. Letter states reason for recall, health risk and action to take; Cease Further Distribution Or Use Of Any Remaining Product Immediately . Distributors are responsible to notify all your customers who may have received this product. Please provide shipment history. Call for questions or additional information: Regulatory 1-860-442-4914 x111 Customer Service 1-860-442-4914 x110 .
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0798-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.