Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile. Recall

Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0808-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 12, 2024
FDA report date
Jan 15, 2025
Quantity in commerce
479 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.

Affected model & UDI checker

Search the exact identifiers published for Z-0808-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CAT1785Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.Check source noticeNot specified
CAT500Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.Check source noticeNot specified
Not specifiedAerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.Check source notice08886479300269
FDA code information

Lot #10235, exp. 2026-05-01; UDI (01)08886479300269(17)260501(10)10235

View the original FDA recall notice

Why was this product recalled?

Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.

What actions did Aerin Medical, Inc. take?

The recalling firm issued emails on 11/12/2024. It was necessary to notify several consignees by mail with an "URGENT: MEDICAL DEVICE FIELD NOTICE" letter dated 11/20/2024. The email described the product purchased by the consignee and the reason for recall. The consignee is informed the programmed treatment time will be shorter than the recommended 30 second treatment time for VivAer and the use of the device may result in insufficient treatment. Under the heading "Action to be Taken," it says if a defective device is identified or the consignee wishes to return unused devices within this production lot, they are to return them to the recalling firm, and return instructions are provided. Regarding the mailed "URGENT: MEDICAL DEVICE FIELD NOTICE" letters dated 11/20/2024, they informed the consignee that some of the units in the production lot were programmed incorrectly and, in these cases, the programmed treatment time will be 12 seconds of radiofrequency energy delivery with no cooling time, not the 30 second treatment time for VivAer that includes 18 seconds of energy delivery and 12 seconds of cooling time. The letter also contains instructions to share this notification with all device users within their facility and network to ensure they are aware of this issue. This notification should also be shared with any organization where the potentially affected devices have been transferred. If you need any further information or support concerning this notification, please contact us at customerservice@aerinmedical.com or call 512-221-9956.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology