Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0809-2025

Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data. Recall

Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0809-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 25, 2024
FDA report date
Jan 15, 2025
Quantity in commerce
2265
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

Affected model & UDI checker

Search the exact identifiers published for Z-0809-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
96280Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.Check source noticeNot specified
Not specifiedXhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.Check source notice10841522107177
Not specifiedXhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.Check source notice10841522100246
FDA code information

Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246

View the original FDA recall notice

Why was this product recalled?

Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.

What actions did Spacelabs Healthcare, Inc. take?

On 11/25/2024, correction notices were sent to customers asking them to do the following: 1) Service manual has been updated by adding a section for "Required periodic system maintenance of Spacelabs Telemetry Farm", pages 5-30 through 5-35. 2) Download current versions of the Service Manual and PM Instructions at https://manuals.spacelabshealthcare.com/ 3) You may also obtain a copy of the latest manual by contacting the firm's Technical Support. 4) Discard previous copies of these documents 5) Complete the online customer acknowledgement form which can be accessed by using URL: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-windows-web-proxy-issue/ For additional information or to report any adverse events as a result of this issue, please contact the firm's Technical Support 1-800-522-7025 and select 2 for Technical Support

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology