Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty. Recall
Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0846-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 18, 2024
- FDA report date
- Jan 22, 2025
- Quantity in commerce
- 24 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty.
Affected model & UDI checker
Search the exact identifiers published for Z-0846-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MRUE200 | Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty. | Check source notice | Not specified |
| Not specified | Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty. | Check source notice | 03700434022926 |
FDA code information
UDI/DI 03700434022926, Lot Numbers: 22A796, 22A797, 22A978, and 22A979
Why was this product recalled?
The Pin Guide is being recalled because the metal tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset being unable to detect the Pin Guide and potentially prolonging surgery time or changing of surgical method from guided to conventional.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Tornier S.A.S. take?
Stryker issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 12/18/2024 via email. The notice explained the problem with the device and requested the following actions be taken: "1. Review the product issue as communicated in this Recall Notice & communicate the issue as appropriate in your organization. 2. Locate and isolate/quarantine the affected device to prevent accidental use pending return. 3. Please use a device from a conforming lot for all procedures. 4. Return the enclosed business reply form by email to confirm receipt of this notification. a. Response is required, even if you may not have any physical inventory on site anymore. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore, please complete even if you no longer have any of the subject devices in your physical inventory. 5. Please ensure that all potential users of these products within your facility are made aware of this notification 6. Please complete the attached response form and return it to memphis.fieldaction@stryker.com so we may arrange for the return of the affected unit(s). 7. Maintain awareness of this communication internally until all required actions have been completed within your facility." Actions for distributors are as follows: "8. If you have further distributed the affected product, please notify your customers at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers. 9. Please inform us of any adverse event and/or report them to the Health/Competent Authorities in accordance with c
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0846-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology