IntelliVue Patient Monitor MX450. Product Number: 866062. Recall
IntelliVue Patient Monitor MX450. Product Number: 866062.
FDA recall Class II
Ongoing
FDA lists this Class II recall for IntelliVue Patient Monitor MX450. Product Number: 866062.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0863-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 31, 2025
- FDA report date
- Dec 10, 2025
- Quantity in commerce
- 1,913,441 units
- Distribution
- See key details
- FDA cause
- Use error
What is being recalled?
IntelliVue Patient Monitor MX450. Product Number: 866062.
Affected model & UDI checker
Search the exact identifiers published for Z-0863-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | IntelliVue Patient Monitor MX450. Product Number: 866062. | Check source notice | 00884838038769 |
FDA code information
Product Number: 866062; UDI-DI: 00884838038769; All Serial No.
Why was this product recalled?
Potential issue where the IntelliVue monitors did not alarm.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips North America Llc take?
An URGENT Medical Device Correction notice, dated 10/31/25, was mailed to customers and distributors. Distributors were notified of this correction and instructed to forward the provided recall notification to all consignees. Additionally, distributors are asked to include their contact information for consignees in the provided response form and perform a good faith effort for each customer to obtain a completed response form. Consignees are instructed to notify all users of this recall notification and forward the notice to all who need to be aware within their organization or where affected devices were potentially transferred. Consignees are to re-evaluate and reconfigure device configuration per the provided instructions. Once recall actions have been completed, consignees are to complete the provided response form either via QR code, by email to Recall.Response@philips.com, or by fax to 877-499-7223. Philips is planning to develop a software update which will enforce the additional confirmation step when the user enables infinite Alarms Off for IntelliVue MP5 models only. Consignees with any questions can contact Philips at 800-722-9377.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0863-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology