Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0871-2026

Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S Recall

Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S

FDA recall Class II

Ongoing

FDA lists this Class II recall for Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0871-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 13, 2025
FDA report date
Dec 10, 2025
Quantity in commerce
20
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S

Affected model & UDI checker

Search the exact identifiers published for Z-0871-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
P36-128-130R-SPhantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-SCheck source noticeNot specified
Not specifiedPhantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-SCheck source notice00889795133670
FDA code information

UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030

View the original FDA recall notice

Why was this product recalled?

Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Paragon 28, Inc. take?

On 10/13/2025, recall notices were emailed to distributors who were asked to do the following: 1) Return affected devices to the firm using return label. 2) Ensure that in instances where you have distributed this affected product, that you forward this Recall Notification to your customers. 3) Replacements are available by contacting Orders@paragon28.com Complaints can be directed to complaints@paragon28.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology