Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0878-2026

MAMMOMAT Inspiration; Recall

MAMMOMAT Inspiration;

FDA recall Class II

Ongoing

FDA lists this Class II recall for MAMMOMAT Inspiration;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0878-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 24, 2025
FDA report date
Dec 10, 2025
Quantity in commerce
6 units (3 US, 3 OUS)
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

MAMMOMAT Inspiration;

Affected model & UDI checker

Search the exact identifiers published for Z-0878-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
10140000MAMMOMAT Inspiration;Check source noticeNot specified
Not specifiedMAMMOMAT Inspiration;Check source notice04056869009063
FDA code information

Model Number: 10140000; UDI-DI: 04056869009063; System Serial Numbers: 6734, 6694, 10121, 10309, 10336, 10342;

View the original FDA recall notice

Why was this product recalled?

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

What actions did Siemens Medical Solutions USA, Inc. take?

On October 24, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Please ensure that all users of the affected products within your organization and others who may need to be informed will receive the relevant safety information provided within this notice and will comply with the recommendations therein. We appreciate your understanding and cooperation with this safety advisory and ask you to immediately instruct your personnel accordingly. Please ensure that this safety advisory is retained in your product related records appropriately. Please keep this information until the corrective measures have been finalized. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification. If this device/equipment is no longer in your possession, please forward this Customer Safety Advisory Notice to the new owner of this device. If applicable, please inform us about the new owner of the device.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history