FDA RECALL RECORD · Z-0887-2025

Breathe Technologies, Inc., Life2000 Ventilation System, REF BT-20-0002AP Recall

Breathe Technologies, Inc., Life2000 Ventilation System, REF BT-20-0002AP

FDA recall Class I

Ongoing

FDA lists this Class I recall for Breathe Technologies, Inc., Life2000 Ventilation System, REF BT-20-0002AP. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0887-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Dec 20, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
1 unit
Distribution
See key details
FDA cause
Process control

What is being recalled?

Breathe Technologies, Inc., Life2000 Ventilation System, REF BT-20-0002AP

Affected model & UDI checker

Search the exact identifiers published for Z-0887-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
BT-20-0002APBreathe Technologies, Inc., Life2000 Ventilation System, REF BT-20-0002APCheck source noticeNot specified
Not specifiedBreathe Technologies, Inc., Life2000 Ventilation System, REF BT-20-0002APCheck source notice00887761978089
FDA code information

UDI/DI 00887761978089 or 00815410020537, Serial Numbers: 122010001081

View the original FDA recall notice

Why was this product recalled?

There is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.

What actions did Baxter Healthcare Corporation take?

Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/20/2024 via USPS first class mail. The notice explained the problem with the device, hazard involved, and requested the following actions be taken: 1. Always have an alternate means of ventilation or oxygen therapy available. 2. Patients may continue using the Life2000 ventilator if it is functioning correctly. If the ventilator becomes inoperable, place the patient on an alternate means of ventilation, if necessary, and contact Baxter Advanced Respiratory, Clinical Support team at 800-397-9071. 3. In the interim if patients experience this failure and have an additional charger available, connect it to the ventilator and verify that the ventilator is charging. 4. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. Log in to the portal using the account number listed in the enclosed reply form instruction sheet. 6. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to the Director of Respiratory Therapy, Patient Safety, and any other departments within your institution who use the affected product. 7. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal. Durable Medical Equipment providers are required to notify patients of this Urgent Medical Device Correction by forwarding the attached home patient letter. For general questions regarding this Urgent Medical Device Correction, please contact the Baxter Advanced Respiratory team at 800

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.