Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Recall

Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type).. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0905-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Dec 18, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
30,140 units
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.

Affected model & UDI checker

Search the exact identifiers published for Z-0905-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MAJ-891Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type).Check source noticeNot specified
Not specifiedOlympus MAJ-891 Forceps/Irrigation Plug (Isolated Type).Check source notice04953170063114
FDA code information

Model Number: MAJ-891. UDI-DI: 04953170063114. All lot numbers.

View the original FDA recall notice

Why was this product recalled?

Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Olympus Corporation of the Americas take?

Olympus notified consignees on about 12/18/2024 via letter. Consignees were instructed to utilize alternative devices to the MAJ-891 according to the provided compatibility table provided, ensure all personnel are completely knowledgeable and thoroughly trained on the content of this letter and the MAJ-891 reprocessing instructions, acknowledge receipt of the letter through the Olympus portal, https://olympusamerica.com/recall, and notify customers if the units have been further distributed. Additionally, Olympus will contact customers to schedule a mutually beneficial time for an Olympus Field Representative to visit each facility and review the contents of this letter and the MAJ-891 reprocessing instructions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.