AMSCO 7052HP Washer/Disinfector Recall
AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
FDA recall Class II
Ongoing
FDA lists this Class II recall for AMSCO 7052HP Washer/Disinfector. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0909-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 19, 2025
- FDA report date
- Dec 17, 2025
- Quantity in commerce
- 10
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
Affected model & UDI checker
Search the exact identifiers published for Z-0909-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 7052HP | AMSCO 7052HP Washer/Disinfector | Check source notice | Not specified |
| Not specified | AMSCO 7052HP Washer/Disinfector | Check source notice | 00724995176587 |
| Not specified | AMSCO 7052HP Washer/Disinfector | Check source notice | 00724995176594 |
| Not specified | AMSCO 7052HP Washer/Disinfector | Check source notice | 00724995176570 |
FDA code information
Lot Code: UDI:00724995176587, 00724995176594, 00724995176570 Serial number of distributed units: 3619625011, 3621625009, 3621625012, 3621925006, 3621925009, 3622425006, 3622725007, 3623425005, 3623825002, 3624525002
Why was this product recalled?
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did STERIS CORPORATION take?
On November 19, 2025, firm began notifying customers via Urgent Medical Device Field Correction letters. All affected customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their units to perform the correction. If the affected unit has not been installed yet, the correction will be performed as part of the installation process. Users can continue to use their units prior to completion of the correction. Under normal operation, the system is designed to detect such electrical arcing specifically, indicating a Drying Over Temperature Switch Tripped alarm. Should your unit alarm, abort the cycle, turn off power to the unit, and contact STERIS.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0909-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology