Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0910-2026

AMSCO 7053HP Washer/Disinfector Recall

AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.

FDA recall Class II

Ongoing

FDA lists this Class II recall for AMSCO 7053HP Washer/Disinfector. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0910-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 19, 2025
FDA report date
Dec 17, 2025
Quantity in commerce
22
Distribution
See key details
FDA cause
Process control

What is being recalled?

AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.

Affected model & UDI checker

Search the exact identifiers published for Z-0910-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
7053HPAMSCO 7053HP Washer/DisinfectorCheck source noticeNot specified
Not specifiedAMSCO 7053HP Washer/DisinfectorCheck source notice00724995176662
Not specifiedAMSCO 7053HP Washer/DisinfectorCheck source notice00724995176679
Not specifiedAMSCO 7053HP Washer/DisinfectorCheck source notice00724995176655
FDA code information

Lot Code: UDI: 00724995176662, 00724995176679, 00724995176655 Serial numbers of distributed units: 3619725008, 3619725011, 3619725013, 3621625010, 3621625011, 3621625013, 3621625014, 3621625015, 3621925001, 3621925002, 3621925005, 3621925010, 3622425005, 3622725005, 3622725011, 3622725012, 3623125006, 3623425016, 3623825008, 3624025002, 3624025008, 3624525007

View the original FDA recall notice

Why was this product recalled?

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did STERIS CORPORATION take?

On November 19, 2025, firm began notifying customers via Urgent Medical Device Field Correction letters. All affected customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their units to perform the correction. If the affected unit has not been installed yet, the correction will be performed as part of the installation process. Users can continue to use their units prior to completion of the correction. Under normal operation, the system is designed to detect such electrical arcing specifically, indicating a Drying Over Temperature Switch Tripped alarm. Should your unit alarm, abort the cycle, turn off power to the unit, and contact STERIS.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology