Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0916-2026

Cardiosave Hybrid. Intra-Aortic Balloon Pump system. Recall

Cardiosave Hybrid. Intra-Aortic Balloon Pump system.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Cardiosave Hybrid. Intra-Aortic Balloon Pump system.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0916-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 4, 2025
FDA report date
Dec 17, 2025
Quantity in commerce
11,470 units
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

Cardiosave Hybrid. Intra-Aortic Balloon Pump system.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Model No. (UDI): D998-00-0800-31 (10607567109053), D998-00-0800-32 (10607567111117), D998-00-0800-33 (10607567109008), D998-00-0800-35 (10607567109107), D998-00-0800-36 (10607567114187), D998-00-0800-45 (10607567108421), D998-00-0800-52 (10607567108438), D998-00-0800-53 (10607567108391), D998-00-0800-55 (10607567108414), D998-00-0800-65 (10607567113432), D998-UC-0800-31 (10607567109053), D998-UC-0800-33 (10607567109008), D998-UC-0800-52 (10607567108438), D998-UC-0800-53 (10607567108391), D998-UC-0800-55 (10607567108414); ALL SERIAL NO.

View the original FDA recall notice

Why was this product recalled?

The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.

What actions did Datascope Corp. take?

An URGENT MEDICAL DEVICE CORRECTION NOTICE, dated October 2025, was sent via FEDEX to consignees. The notification notifies consignees of four addendum updates: 1) NVRAM Preventive Maintenance (New FSCA - reflected in this recall entry), 2) Vibration and Shock Table Specifications (New FSCA), 3) Helium O-Ring Preventive Maintenance (FSCA 2249723-01/24/2023-004-C), and 4) Battery Runtime Specifications (FSCA 2249723-09/10/2021-001-R). The notification instructs consignees regarding the NVRAM Preventive Maintenance to review the provided IFU addendum, update local preventive maintenance procedures as such, and ensure all NVRAM batteries are within their life expectancy. If a NVRAM is greater than 8 years old, until it can be replaced, it is advised a back-up Cardiosave IABP be available to provide therapy. Consignees with any questions can contact Datascope/Getinge Customer Support at 1-88-943-8872 (options 4, 2, 1) Monday through Friday, from 8:00 AM EST and 6:00 PM EST.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology