Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0919-2026

Stryker Arise 1000EX mattress, Part Number 2236000000 Recall

Stryker Arise 1000EX mattress, Part Number 2236000000

FDA recall Class II

Ongoing

FDA lists this Class II recall for Stryker Arise 1000EX mattress, Part Number 2236000000. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0919-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 31, 2025
FDA report date
Dec 17, 2025
Quantity in commerce
297 units
Distribution
See key details
FDA cause
Release of Material/Component prior to receiving test results

What is being recalled?

Stryker Arise 1000EX mattress, Part Number 2236000000

Affected model & UDI checker

Search the exact identifiers published for Z-0919-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedStryker Arise 1000EX mattress, Part Number 2236000000Check source notice00857268006608
FDA code information

GTIN 00857268006608, Serial Numbers: 1000502893, 1000281054, 1000281055, 1000281056, 1000213521, 1000213522, 1000213529, 1000213530, 1000213531, 1000213532, 1000213533, 1000214321, 1000214331, 1000377179, 1000377180, 1000377181, 1000382290, 1000382291, 1000382292, 1000382293, 20090709143150, 1000537621, 1000537622, 1000481695, 1000481811, 1000502644, 1000502645, 1000469456, 1000469457, 1000469821, 1000469822, 1000065720, 1000066888, 1000502643, 1000502895, 1000508702, 1000510080, 1000510081, 1000510082, 1000510098, 1000183920, 1000183921, 1000184053, 1000451149, 1000451347, 1000485570, 1000232648, 1000232649, 1000232650, 1000232651, 1000232652, 1000512470, 1000514147, 1000510095, 1000510096, 1000482548, CU400-1983BA355214, 1000452867, 1000452869, 1000475796, 1000475797, 1000475867, 1000475868, 1000485537, 1000485571, 1000485538, 1000471627, 1000471860, 3100000471, 3100000874, 1000486159, 1000495054, 1000495055, 1000484745, 1000484746, 1000484747, 1000456813, 1000467673, 1000483228, 1000483229, 1000483465, 1000483466, 1000483467, 1000483468, 1000468975, 1000468976, 1000483230, 1000514148, 1000514149, 1000455365, 1000455367, 1000456815, 1000053839, 1000181244, 1000482317, 1000482318, 1000482319, 1000276077, 1000393550, 1000481813, 1000370748, 1000372689, 1000372690, 1000372691, 1000372692, 1000372693, 1000372694, 1000372695, CU400-1375, CU400-1447, 1000451346, 1000361096, 1000361097, 1000422407, 1000422427, 1000422428, 1000484748, 1000502891, 1000502892, 1000535388, 1000160222, 1000500377, 1000500378, 1000500654, 1000504503, 1000504504, 1000504505, 1000055394, 1000260702, 1000260703, 1000260704, 1000454956, 1000435830, 1000435831, 1000514150, 1000446907, 1000446908, 1000335539, 1000411780, 1000417224, 1000007185, 1000007187, 1000007188, 1000007189, 1000046411, 1000046735, 1000048762, 1000512471, 1000512474, 1000514151, 1000514152, 1000514153, 1000514154, 1000514155, 1000521760, MX110241, MX110245, 1000436370, 1000436372, 1000482545, 1000482546, 1000482547, 1000454955, 1000454957, 1000455328, 1000455329, 1000387606, 1000387836, 1000480162, 1000373921, 1000373922, 1000373923, 1000373924, 1000389360, 1000389361, 1000500655, 1000431889, 1000431891, 1000431892, 1000431893, 1000431894, 1000446831, 1000481812, 1000476944, 1000476945, 1000479698, 1000479699, 1000479700, 1000480160, 1000480161, 1000483231, 1000483994, 1000483995, 1000483996, 1000483997, 3100000278, 1000354259, 1000413269, 1000413270, 1000413271, 1000413272, 1000417222, 1000417223, 1000418007, 1000418008, 1000418009, 1000422044, 1000422045, 1000422046, 1000422047, 1000422098, 1000422099, 1000422100, 1000422101, 1000422102, 1000422103, 1000422104, 1000422105, 1000422106, 1000422107, 1000422108, 1000422109, 1000422404, 1000422405, 1000422406, 1000422429, 1000422430, 1000422431, 1000422432, 1000422433, 1000422434, 1000426512, 1000426513, 1000426514, 1000426515, 1000426516, 1000426517, 1000426518, 1000426519, 1000426795, 1000426796, 1000426797, 1000426798, 1000426799, 1000426800, 1000426801, 1000426802, 1000427341, 1000427342, 1000427343, 1000427344, 1000427345, 1000427346, 1000427347, 1000427348, 1000427349, 1000427350, 1000427351, 1000427352, 1000427353, 1000427354, 1000427355, 1000427356, 1000427634, 1000427635, 1000427636, 1000427637, 1000427638, 1000427639, 1000427640, 1000427641, 1000427701, 1000427702, 1000427703, 1000427704, 1000427705, 1000427706, 1000427707, 1000427708, 1000428096, 1000428097, 1000428098, 1000428099, 1000428100, 1000428101, 1000428102, 1000428103, 1000428297, 1000232647, 1000531402, 1000531401, 1000531403, 1000524879, 1000524877, 1000524878, 1000531400, 1000524880.

View the original FDA recall notice

Why was this product recalled?

Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Stryker Medical Division of Stryker Corporation take?

Stryker issued an URGENT: VOLUNTARY MEDICAL DEVICE SAFETY notice to its sales representatives on 11/13/2025 for hand delivery to its consignees on 11/14/2025. The notice explained the issue, potential hazard, and requested the following: "Actions needed: 1. Locate the MV3 and Arise 1000EX devices that have been shipped to your facility. Serial numbers are provided on the enclosed business reply form. 2. If an MV3 is being used together with an Arise 1000EX mattress, Stryker recommends the use to be discontinued upon receipt of this notification. 3. Complete the enclosed business reply form to confirm receipt of this notification. 4. Return your completed business reply form to Stryker by fax +1 269 488 8691 or email productfieldaction@stryker.com. 5. For those requesting a replacement of a compatible mattress, Stryker will contact you upon receipt of your Business Reply Form for next steps. 6. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location." For questions or concerns, contact Customer Service at +1 800 327 0770, Option 1 for Customer Support, Monday through Friday from 8:00 a.m. to 7:00 p.m. ET.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.