Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0933-2025

ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractors Recall

ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractors

FDA recall Class II

Ongoing

FDA lists this Class II recall for ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractors. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0933-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 3, 2024
FDA report date
Jan 22, 2025
Quantity in commerce
1050 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractors

Affected model & UDI checker

Search the exact identifiers published for Z-0933-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
96-2243ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractorsCheck source noticeNot specified
96-2243MECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractorsCheck source noticeNot specified
96-2267ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractorsCheck source noticeNot specified
96-2267AECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractorsCheck source noticeNot specified
96-2269AECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractorsCheck source noticeNot specified
96-2959ECONO STERILE medical procedure kits labeled as: 1) ECONO STERILE ARMY RETR (SET OF 2) STER CS25, Model Number 96-2243; 2) ECONO STERILE ARMY RETR (SET OF 2)STRL CS/25, Model Number 96-2243M; 3) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267; 4) ECONO STERILE SENN RETR DE BL 6.25"STRLCS/25, Model Number 96-2267A; 5) ECONO STERILE SENN RETR DE SH 6.25"STRLCS/25, Model Number 96-2269A; 6) ECONO STERILE MINI HOHMANN 6MM 6.5" STERCS25, Model Number 96-2959; retractorsCheck source noticeNot specified
FDA code information

1) 96-2243, UDI/DI 50649111493087Lot codes: SK-115, SK-116, SK-129; 2) 96-2243M, UDI/DI 50649111448605Lot codes: SK-129; 3) 96-2267, UDI/DI 50649111448698Lot codes: SK-117; 4) 96-2267A, UDI/DI 50649111487604Lot codes: SK-112; 5) 96-2269A, UDI/DI 50649111487611Lot codes: SK-112; 6) 96-2959, UDI/DI 50649111493070Lot codes: SK-118.

View the original FDA recall notice

Why was this product recalled?

Reports of various packaging issues that may result in a breach of the sterile barrier.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Sklar Instruments take?

On December 3, 2024 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Action required: 1. Check your inventory for any affected product lots (see Attachment A). If you have affected product in inventory, discontinue use immediately. 2. Notify Additional Users. If you have further distributed any of the items referenced in this notification, you must notify those downstream customers of this recall. Forward a copy of this letter, along with copies of Attachment A and the Recall Acknowledgement form, for such customers to complete and return to Sklar. 3. Notify Sklar. Please complete the enclosed Recall Acknowledgement form as soon as possible, indicating whether you do or do not have any of the affected products remaining in your inventory and whether you have distributed any of the affected products to downstream customers. Return the form to Sklar by e-mail to surgi@sklarcorp.com. 4. Return the affected product. If you do have any of the affected products in inventory, Sklar customer service will contact you to arrange for the affected product to be returned. Sklar will credit its direct customers for returned affected product once it receives the affected product back. Sklar Instruments is committed to providing high quality, safe, and effective products. We sincerely apologize for any inconvenience this action may cause your operations. If you have experienced an adverse event associated with any of these products, please contact us and/or report the event to FDA s MedWatch Adverse Event Reporting program online.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology