Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0936-2026

5008X CAREsystem +CLiC +CDX; Recall

5008X CAREsystem +CLiC +CDX;

FDA recall Class II

Ongoing

FDA lists this Class II recall for 5008X CAREsystem +CLiC +CDX;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0936-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 25, 2025
FDA report date
Dec 24, 2025
Quantity in commerce
115 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

5008X CAREsystem +CLiC +CDX;

Affected model & UDI checker

Search the exact identifiers published for Z-0936-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
M2044415008X CAREsystem +CLiC +CDX;Check source noticeNot specified
Not specified5008X CAREsystem +CLiC +CDX;Check source notice00840861102433
FDA code information

Model Number: M204441; UDI-DI: 00840861102433; All serial numbers produced with a software version prior to 4.82.4;

View the original FDA recall notice

Why was this product recalled?

Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Fresenius Medical Care Holdings, Inc. take?

On Octiber 25, 2025, customers were called and read a Medical Device Recall Script. Instructions: Please examine the 5008X CAREsystems at your location to confirm that you have any of this product: 1. Please segregate this product until the software update can be performed. 2. Please provide the number of 5008X CAREsystems you have installed at your clinic and the serial numbers of the devices. If the customer has any additional medical concerns or questions, please have them contact Medical Information and Communication: 855-616-2309 or Website: www.freseniusmedinfo.com. Report any complaints or adverse events to product.complaints@fmc-na.com or FDA MedWatch at https://www.fda.gov/safety/medical-product-safety-information/medwatchforms-fda-safety-reporting

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology