muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate Recall
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
FDA recall Class II
Completed
FDA lists this Class II recall for muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0956-2025
- Recall classification
- Class II
- Enforcement status
- Completed
- Initiation date
- Jun 21, 2021
- FDA report date
- Jan 22, 2025
- Quantity in commerce
- 275 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
Affected model & UDI checker
Search the exact identifiers published for Z-0956-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 2050 | muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate | Check source notice | Not specified |
FDA code information
Soft ware version 1.49.02 UDI: +B595BSM20500/\$\$7V1.49.02/16D20200504S and Software version 1.52.00 UDI: +B595BSM20500/\$\$7V1.52.00/16D20210305K
Why was this product recalled?
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
What actions did Murata Vios, Inc. take?
Murata Vios notified their consignees via email on 06/21/2021. The notice explained the issue and provided workflow guidance to prevent the problem from occurring while a software update was being created. The firm followed up with another email on 08/24/2021 informing the consignees that the software update was ready for deployment and to request scheduling an appointment for the upgrade.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0956-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026