Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0965-2026

LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02 Recall

LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02

FDA recall Class II

Ongoing

FDA lists this Class II recall for LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0965-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 24, 2025
FDA report date
Dec 31, 2025
Quantity in commerce
7 customers
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02

Affected model & UDI checker

Search the exact identifiers published for Z-0965-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedLifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02Check source notice00887709127227
FDA code information

version 2.2.1 UDI-DI 00887709127227 Issue affects customers who migrated from LifeShield DLM v2.1x to LifeShield DLM v2.2.1 and have any version of a Plum Duo v1.1 drug library that was approved or locked for review before the migration.

View the original FDA recall notice

Why was this product recalled?

For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.

What actions did Icu Medical, Inc. take?

Firm began notifying customers on 11/24/2025 via Urgent Medical Device Correction letters. Customers were instructed to ensure all users or potential users of LifeShield ISSS DLM 2.2.1 with Plum Duo 1.1 drug libraries are made aware of the issue and the following proposed mitigations: Until a software update is released, when performing the first Plum Duo v1.1 drug library review after the migration, the pharmacist is able to review the changes using alternate review mechanisms, as noted in the LifeShield Drug Library Management User Manual: a) Accessing the version history view of the drug library b) Review drug library reports in CSV or HTML format c) Assess each change within LifeShield DLM to verify the accuracy of the changes (i.e., view impacted medication ruleset hard dose rate limit or view a CCA to verify a maximum rate setting) The firm is developing a software update to correct the issue and will notify customers of the timing of the software update.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology