Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0976-2025

BD Pyxis Med 4000 Auxiliary (AUX), REF 314 Recall

BD Pyxis Med 4000 Auxiliary (AUX), REF 314

FDA recall Class II

Ongoing

FDA lists this Class II recall for BD Pyxis Med 4000 Auxiliary (AUX), REF 314. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0976-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 14, 2023
FDA report date
Jan 29, 2025
Quantity in commerce
5 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

BD Pyxis Med 4000 Auxiliary (AUX), REF 314

Affected model & UDI checker

Search the exact identifiers published for Z-0976-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
314BD Pyxis Med 4000 Auxiliary (AUX), REF 314Check source noticeNot specified
Not specifiedBD Pyxis Med 4000 Auxiliary (AUX), REF 314Check source notice10885403512650
FDA code information

UDI/DI 10885403512650, Serial Numbers: 16140589, 16194078, 16194079, 16194081, 16079165.

View the original FDA recall notice

Why was this product recalled?

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

What actions did CareFusion 303, Inc. take?

BD issued an Urgent Medical Device Product Advisory to its consignees on 02/14/2023 via letter. The notice explained the problems with the device, health hazards, and requested the following actions be taken: "1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them. 3. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. 4. If you have specific questions about this letter, please contact 1-866-583-8783." The firm issued a second notice on 01/09/2024 with the same information.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology