stryker 3.0MM Precision Neuro Match Head, REF 5820-107-530, used to cut bone and bone cement during neurosurgery Recall

stryker 3.0MM Precision Neuro Match Head, REF 5820-107-530, used to cut bone and bone cement during neurosurgery

FDA recall Class II

Ongoing

FDA lists this Class II recall for stryker 3.0MM Precision Neuro Match Head, REF 5820-107-530, used to cut bone and bone cement during neurosurgery. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0982-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 28, 2023
FDA report date
Jan 29, 2025
Quantity in commerce
1489 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

stryker 3.0MM Precision Neuro Match Head, REF 5820-107-530, used to cut bone and bone cement during neurosurgery

Affected model & UDI checker

Search the exact identifiers published for Z-0982-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5820-107-530stryker 3.0MM Precision Neuro Match Head, REF 5820-107-530, used to cut bone and bone cement during neurosurgeryCheck source noticeNot specified
Not specifiedstryker 3.0MM Precision Neuro Match Head, REF 5820-107-530, used to cut bone and bone cement during neurosurgeryCheck source notice07613154028177
FDA code information

GTIN 07613154028177, Lot Numbers: 22179047, 22179027, 22160017

View the original FDA recall notice

Why was this product recalled?

There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-430 but package contains a 5820-107-530 device or vice versa).

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Stryker Corporation take?

Stryker issued a Medical Device Recall notice to its consignees on 06/28/2024 via FedEx. The notice explained the issue and requested the following: "Actions to be taken 1. Review your inventory to locate and quarantine any affected products found on-hand. 2. Sign and return the enclosed Business Response Form (BRF), even if affected product is no longer in inventory, to the email Instruments.recalls@stryker.com. 3. Upon receiving your Business Response Form to the email, Instruments.recalls@stryker.com, a shipping label will be provided to return any recalled product on-hand. 4. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. Please contact Jordan Anderson at (269) 800-1347 or Instruments.recalls@stryker.com with questions or concerns."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.