Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0992-2025

Medline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690 Recall

Medline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690

FDA recall Class I

Ongoing

FDA lists this Class I recall for Medline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0992-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Dec 31, 2024
FDA report date
Feb 5, 2025
Quantity in commerce
58 kits
Distribution
See key details
FDA cause
Process design

What is being recalled?

Medline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690

Affected model & UDI checker

Search the exact identifiers published for Z-0992-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690Check source notice10653160312727
Not specifiedMedline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690Check source notice00653160312720
Not specifiedMedline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690Check source notice10653160318019
Not specifiedMedline General Surgery Tray kits: 1) ARTERIAL LINE TRAY, Item Number ART600; 2) A LINE TRAY, Item Number ART690Check source notice00653160318012
FDA code information

1) ART600 UDI-DI: 10653160312727 (each), 00653160312720 (case), Lot Number 2024110190; 2) ART690 UDI-DI: 10653160318019 (each), 00653160318012 (case), Lot Number 2024092490

View the original FDA recall notice

Why was this product recalled?

The catheter subassembly in the Integrated Arterial Catheter was manufactured with excess material on the catheter hub.

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

A MEDICAL DEVICE RECALL notification letter dated 12/31/24 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-24-278-FGX1 Recall Code: RECALL CODE 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology