Rsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures. Recall
R¿sch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The R¿sch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Rsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-0995-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 26, 2025
- FDA report date
- Dec 31, 2025
- Quantity in commerce
- 23 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Rsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
Affected model & UDI checker
Search the exact identifiers published for Z-0995-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Rsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures. | Check source notice | 00827002069299 |
FDA code information
Reference Part Number RUPS-100: Order Number G06929, UDI (01)00827002069299(17)280624(10)16659065, Lot Number 16659065
Why was this product recalled?
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
What actions did Cook Incorporated take?
Cook Medical issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 11/26/2025 by email and/or letter delivered via a courier. The notice explained the issue, risk to health, and requested the identification and return of the affected product. The notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. For questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0995-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology