Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0998-2026

The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement. Recall

The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.

FDA recall Class II

Ongoing

FDA lists this Class II recall for The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0998-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 27, 2025
FDA report date
Jan 14, 2026
Quantity in commerce
56
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Part Number V000629

View the original FDA recall notice

Why was this product recalled?

Vision RT is informing customers of an omission identified in the Instructions for Use for AlignRT InBore systems. Information identifying the lasers and their characteristics was omitted from the Instructions for Use for the subject system as required by 21 CFR 1040,10(h), via conformance with recognized consensus standard IEC 60825-1 Ed. 3 (FDA Laser Notice 56).

What actions did Vision RT Ltd take?

Vision RT wrote to inform customers of an omission that they have identified in the Instructions for Use for AlignRT InBore systems. This notification contained information for users of AlignRT InBore systems on the content omitted and how to access the corrected Instructions for Use.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA notice posted

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology