Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0999-2026

LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471 Recall

LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471

FDA recall Class II

Ongoing

FDA lists this Class II recall for LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0999-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 5, 2025
FDA report date
Jan 14, 2026
Quantity in commerce
U.S.: 1, OUS: 95
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471

Affected model & UDI checker

Search the exact identifiers published for Z-0999-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11574100LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471Check source noticeNot specified
10762471LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471Check source noticeNot specified
FDA code information

(01)04056869269931(21)10330

View the original FDA recall notice

Why was this product recalled?

A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Medical Solutions USA, Inc. take?

The Customer Advisory Notice (CAN) was sent to the affected customer on December 5, 2025. The CAN notifies the customer of the problem and provides steps the user should take to avoid the problem until the resolution software update is available. When switching within five seconds from a table related OGP to a bucky wall stand related OGP with top-alignment and then back to a table related OGP, it may happen, that the collimation is not adapted to the limitations of the detector. The active radiation field may thus extend beyond the limits of the detector, resulting in radiation outside of the area of the detector.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology