Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM, 35CM, Front-actuated Grip Type S Recall
Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM, 35CM, Front-actuated Grip Type S¿ Model Number: TB-0535FCS Catalog Number/Product Code: N5423810 Software Version: N/A Product Description: Thunderbeat instrument is a sterile, single use item which is intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (WSG-400), and the Thunderbeat Transducer, (TD-TB400).
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM, 35CM, Front-actuated Grip Type S. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1003-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 25, 2024
- FDA report date
- Jan 29, 2025
- Quantity in commerce
- 27,352 units (OUS ONLY)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM, 35CM, Front-actuated Grip Type S Model Number: TB-0535FCS Catalog Number/Product Code: N5423810 Software Version: N/A Product Description: Thunderbeat instrument is a sterile, single use item which is intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (WSG-400), and the Thunderbeat Transducer, (TD-TB400).
Affected model & UDI checker
Search the exact identifiers published for Z-1003-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| N5423810 | Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM, 35CM, Front-actuated Grip Type S | Check source notice | 04953170383564 |
| TB-0535FCS | Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM, 35CM, Front-actuated Grip Type S | Check source notice | Not specified |
| Not specified | Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM, 35CM, Front-actuated Grip Type S | Check source notice | 04953170383561 |
FDA code information
Model: TB-0535FCS; Catalog Number/Product Code: N5423810; UDI: 04953170383564, 04953170383561; Lots: All
Why was this product recalled?
The probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
What actions did Olympus Corporation of the Americas take?
On December 20, 2024, URGENT FIELD SAFETY NOTICE: ADVISORY letters were sent to customers. Actions Required: Olympus requires you to take the following actions: 1. Carefully read the content of this FSN. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this FSN and the relevant IFU. 3. Olympus requests that you acknowledge receipt of this FSN through our web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number "0462 c. Complete the form as instructed and include your account ID number: [XXXXXXXX] d. Indicating in the comments if you would like to receive in-service training to support the correct use of this device. 4. If you have further distributed this product, identify your customers, and forward this FSN to them. Olympus requests that you report any complaints, including probe damage or pad detachment, to Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1), and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@olympus.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1003-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology