BD PCR Cartridges used with MAX System. Recall
BD PCR Cartridges used with MAX System. Model Number: 437519
FDA recall Class II
Ongoing
FDA lists this Class II recall for BD PCR Cartridges used with MAX System.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1004-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 19, 2024
- FDA report date
- Jan 29, 2025
- Quantity in commerce
- 516,435 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
BD PCR Cartridges used with MAX System. Model Number: 437519
Affected model & UDI checker
Search the exact identifiers published for Z-1004-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 437519 | BD PCR Cartridges used with MAX System. | Check source notice | Not specified |
| Not specified | BD PCR Cartridges used with MAX System. | Check source notice | 30382904375196 |
FDA code information
Model Number: 437519. UDI Numbers: (01)30382904375196(17)261205(10)4323345, (01)30382904375196(17)261129(10)4317725, (01)30382904375196(17)261122(10)4309022, (01)30382904375196(17)261115(10)4302135, (01)30382904375196(17)261111(10)4296086, (01)30382904375196(17)261104(10)4289593, (01)30382904375196(17)261029(10)4281532, (01)30382904375196(17)261021(10)4275828, (01)30382904375196(17)261014(10)4268747, (01)30382904375196(17)261007(10)4254714, (01)30382904375196(17)260930(10)4254710, (01)30382904375196(17)260923(10)4247089, (01)30382904375196(17)260913(10)4239079, (01)30382904375196(17)260909(10)4233297, (01)30382904375196(17)260902(10)4225364, (01)30382904375196(17)260826(10)4204545, (01)30382904375196(17)260805(10)4198065, (01)30382904375196(17)260729(10)4191581, (01)30382904375196(17)260722(10)4185634, (01)30382904375196(17)260716(10)4178328, (01)30382904375196(17)260704(10)4164798, (01)30382904375196(17)260701(10)4164796, (01)30382904375196(17)260701(10)4164792, (01)30382904375196(17)260701(10)4164794, (01)30382904375196(17)260701(10)4164789, (01)30382904375196(17)260624(10)4149773, (01)30382904375196(17)260617(10)4149768, (01)30382904375196(17)260610(10)4143963, (01)30382904375196(17)260605(10)4142324, (01)30382904375196(17)260604(10)4136790, (01)30382904375196(17)260528(10)4130663,(01)30382904375196(17)260527(10)4130589, (01)30382904375196(17)260520(10)4121669, (01)30382904375196(17)260517(10)4117376, (01)30382904375196(17)260513(10)4114952, (01)30382904375196(17)260506(10)4107497, (01)30382904375196(17)260429(10)4100041, (01)30382904375196(17)260422(10)4093507.
Why was this product recalled?
Potential for signal drift in specific lots of BD PCR Cartridges.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Becton Dickinson & Co. take?
BD notified consignees via email and mail on about 12/19/2024 with a letter titled, URGENT: Medical Device Product Correction. Consignees were informed regarding when they can and cannot use the affected BD PCR Cartridges. They were instructed to inspect inventory for affected cartridges, ensure the recall notification is read and understood by all affected personnel, to provide the notification to all affected personnel and customers if further distributed, to complete and return the Customer Response Form, and to report any adverse events associated with the affected units. Distributors were provided the same information and instructions, but were also instructed to provide a copy of the letter to their customers.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1004-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology